Infitek Microwave Digestion System, MDS-700-T18

Infitek Microwave Digestion System, MDS-700-T18

Optimize your lab's efficiency with the Infitek Microwave Digestion System, MDS-700-T18, featuring:
- 18 vessels with a 100mL volume each
- Seamless operation with:
  - Precise temperature monitoring using a contact-less IR sensor
  - An intuitive 7-inch touch screen display

  • PRODUCT INFO

    It has been widely used in food, textile, geology, metallurgy, plastics, coal, cosmetics, petrochemicals, biomedicine, environmental monitoring, sewage treatment and other fields.

    Features

    • Special designed sample digestion vessel

    The automatic vent and self-resealing structure ensures the digestion vessels can automatically release pressure and instantly reseal when a sudden over-pressure situation occurs.

    • Contact-lesstemperature and pressure monitoring system

    The advanced contact-less IR sensor could measure the real-time temperature of sample solutions in each digestion vessel. Meanwhile, the contact-less pressure sensor could monitor the real-time pressure of each vessel.

    • Large storage capacity

    It can edit and store 255 types of programs according to user’ s requirements and each method program can set parameters (temperature, pressure, time, microwave power) according to their own requirements which provides convenient operation for users.

    • Safety protection system

    With high-strength double locked security door, the real-time temperature and pressure monitoring, automatic adjustment of over-pressure and over-temperature and abnormal sound monitoring, the device can run in a highly safe environment.

    • Power adjustment

    The variable frequency resonance or non-pulse(optional) continues rotating mode ensures high efficient and even microwave throughput. it can be set between 0~2000W/0~3000W according to user’ s requirements.

  • Product Specification

    Model MDS-700-T18
    Vessel Quantity 18
    Vessel Volume 100mL
    Temperature Monitoring System Temperature monitoring: Contactless IR sensor

     

    Temperature control:Scanning control of each vessel

    Temperature control range: 50~400℃;

    Temperature control accuracy:±0.3℃

    Pressure Monitoring System Contact-less pressure sensor

     

    Pressure control range:0~10MPa

    Pressure control accuracy:0.01MPa

    Vessel Material Inner: imported TFM

     

    Outer: imported PEEK+glass fiber

    Display 7-inch touch screen
    Rotation Mode 360° Continuous rotation
    Microwave Power 0-2000W adjustable/ 0-3000W adjustable
    Microwave Leak <5mW/cm²
    Microwave Frequency 2450MHz
    Max. Working Pressure 6MPa
    Max. Working Temperature 250℃
    Inner Vessel Temperature Limit 300℃
    Electricity AC 220V, 16A, 50/60Hz
    Dimension(L*W*H) 640*630*590mm
    N.W./G.W. 75kg/148kg(Main device+Accessories)
    Shipping Dimension(L*W*H) 800*800*810mm (Main device),620*590*560mm (Accessories)