top of page
Infitek Kjeldahl Analyzer, KJA-P500

Infitek Kjeldahl Analyzer, KJA-P500

Kjeldahl Analyzer

 

  • Measuring Range: 0.08~240mgN
  • MeasuredSample Volume: solid < 6g, liquid <16mL
  • Recovery Rate: > 99.5%
  • Repeatability(RSD): ≤ 0.5%
  • Product Info

    Application

    It is widely used to detect the overall content of ammonia nitrogen and protein nitrogen in grain, oil, food, dairy products, beverages, feed, soil, chemical fertilizers, drugs, sediments and chemical products. It is an important physical and chemical analysis instrument for product quality inspection.

    Features

    Accuracy

    • Intermittent alkali addition ensures that the acid-base reaction is controllable to eliminate splashing during sample distillation.
    • The constant pressure design prevents the solution from being sucked back into the collection bottle.

    Safety

    • The distiller adopts double liquid level control (double safety) to prevent dry burning.​

    Convenience

    • One-touch start of operation interface, LCD touch screen, easy to operate and enter the test program with one step.
    • Large LCD screen display, independent circuit, water and gas, external reagent bucket.
    • Alkali can be added manually at any time during the distillation to ensure the accuracy of sample data.
    • Manual and auto mode can be switched freely, and the entire test process can be tracked and displayed in real time.
    • Data storage and editing methods: Automatic preheating, cleaning and acid/alkali/dilution addition, and 999 programs can be stored.

    Beautiful appearance

    • The instrument shell is made of ABS engineering plastic mold, which is resistant to strong acid and alkali, and heat insulated.
    • High temperature, acid and alkali resistant materials are used to ensure the internal anti-corrosion, anti-leakage and sealing inside the instrument.
    • High temperature resistant, acid and alkali resistant materials are used to ensure the internal corrosion resistance, leakage prevention and sealing of the instrument.
  • Specifications

    Model KJA-P500
    Measuring Range 0.08~240mgN
    Measured Sample Volume Solid < 6g, liquid <16mL
    Recovery Rate > 99.5% (1~240mgN)
    Repeatability (RSD) ≤ 0.5% (1~240mgN)
    Sample Testing Time 3-8min
    Distillation Time 1-99min
    Distillation Power 1,500W, adjustable
    Delay Time 0-1,800s (easy to detect fertilizer)
    Distillation Capacity 15-30mL/min, adjustable
    Data Storage 999 distillation programs
    Water Supply Water pressure > 0.15MPa
    Cooling Water Consumption > 1.5L/min (Water temperature ≤ 20°C)
    Shell One-time molding by using ABS engineering plastics
    Function Automatic addition of acid, alkali, dilute solution, automatic replenishment of water, automatic distillation and data storage, etc.
    Electricity AC 220V ± 10%, 50-60Hz
    External Dimension 370*300*680mm
    N.W./G.W. 15kg/26kg
    Package Dimension 495*440*860mm