Ro-Chain RC-AM999 Anaesthesia machine
Experience precise patient care with the Ro-Chain RC-AM999 Anesthesia Machine. This reliable, state-of-the-art machine ensures safe anaesthesia administration, favored by medical professionals. Advanced features and user-friendly controls enhance efficiency and patient outcomes. Trust Winselluma for quality and exceptional service. Elevate your practice with the Ro-Chain RC-AM999 Anesthesia Machine.
Ventilator Specifications
1. Modes of Ventilation
Mode Description VCV Volume Control Ventilation SIMV-V Synchronized Intermittent Mandatory Ventilation – Volume SIMV-P Synchronized Intermittent Mandatory Ventilation – Pressure MANUAL Manual Mode A/C Assist/Control PCV Pressure Control Ventilation PSV Pressure Support Ventilation STAND-BY Stand-by Mode PA-VC (Optional) Pressure-Assisted Volume Control (Optional) 2. Ventilator Parameters
Working Principle
-
Electronic control, pneumatic driven
Performance Parameters
-
Tidal Volume Range: 10–1500 mL
-
Pressure Range: 5–60 cmH₂O
-
Respiration Rate: 1–100 bpm (1–40 bpm in SIMV)
-
I:E Ratio: 4:1 to 1:8
-
Pressure Trigger Range: –20 to +20 cmH₂O (based on PEEP)
-
Flow Trigger: 0–15 L/min
-
Minute Volume: >18 L/min
-
SIGH: 0–5 per 100 breaths
3. System Standards
Safety Systems
-
Hypoxic Guard: N₂O cutoff valve, O₂ concentration >25%
-
Safety Valve: <12.5 kPa
-
Fresh Gas Compensation: 25–75 L/min
CO₂ Absorption
-
Absorber Volume: 2 L
Flowmeter
-
Electronic flowmeter
-
Ranges:
-
O₂: 0.1–15 L/min
-
N₂O: 0.1–15 L/min
-
Air: 0.1–15 L/min
-
Vaporizer
-
2 Selectatec mounts (maximum 2)
-
Agents: Isoflurane / Enflurane / Sevoflurane / Halothane
4. Monitoring
Display & Graphics
-
Waveforms: P-T, F-T, V-T
-
Loops: P-V, V-F, F-P
-
Functions: Zoom in/out, comparison, freeze
Monitored Parameters
-
Tidal volume
-
Minute volume
-
Airway pressure
-
I:E ratio
-
Lung compliance
-
Plateau pressure
-
O₂ concentration
-
CO₂ (optional)
-
Anaesthetic gas monitoring (optional)
5. Alarm Limits
Audio/Visual Alarms for:
-
No tidal volume
-
Minute volume limit
-
Airway pressure limit
-
FiO₂ limit
-
O₂ failure
-
Air failure
-
AC power failure
-
Low battery
6. PEEP
-
Type: Integrated, electronic control
-
Range: OFF, 3–30 cmH₂O
7. Auxiliary Features
-
ACGO: Auxiliary Common Gas Outlet
-
Yoke System: Pin-index; supports 11 L cylinders (O₂ / N₂O)
-
AGSS: Anaesthetic Gas Scavenging System (optional)
-
Communication Interfaces: Ethernet, USB, VGA
-
Worktable: Optional
-
Patient Monitor: Optional
-
E-Anesthesia Record System: Optional
-
Auxiliary Oxygen Meter: Optional
-