Ro-Chain RC-200C CPAP SYSTEM
Experience unparalleled sleep therapy with the Ro-Chain RC-200C CPAP System, designed for superior comfort and effective treatment of sleep apnea and other related medical conditions. As a trusted distributor, winselluma ensures you receive top-tier, reliable products tailored to your healthcare needs. The RC-200C offers advanced technology that guarantees optimal airflow, fostering better sleep and improved overall health. Trust in winselluma's commitment to quality and excellence in providing medical solutions that genuinely enhance your well-being.
Applications
The CPAP system is designed for newborns with strong spontaneous breathing and mild respiratory failure. It is also suitable for use in pediatric clinics to provide NCPAP oxygen therapy, supporting effective respiratory care and oxygenation in infants.
Product Features:
CPAP System – Performance and Features
Feature Description CPAP Ventilation Mode Maintains spontaneous breathing in children while reducing respiratory effort; provides constant flow and limited pressure CPAP for newborns Air/Oxygen Delivery Delivers properly adjusted air/O₂ mixture with continuous positive pressure for premature babies, newborns, and children, enhancing the safety and efficacy of oxygen therapy Oxygen/Air Blender Built-in blender ensures stable and precise O₂ concentration adjustment Oxygen Monitoring High-quality oxygen sensor monitors real-time O₂ concentration and triggers alarms for unsafe levels, ensuring breathing safety CPAP Control Reliable CPAP control valve combined with advanced pressure monitoring improves adjustment precision Display 3" display for clear monitoring of parameters Working Condition
Parameter Specification Air Source Oxygen, Air Air Source Pressure 280 – 600 kPa Power Supply With air compressor: ~220 V ± 22 V, 50 Hz
Without air compressor: 100–240 V, 50/60 HzInput Power With air compressor: 900 VA
Without air compressor: 250 VAVentilation Mode CPAP Specifications
Adjusting Parameters
Parameter Range Inhaled Oxygen Concentration 21% – 100% CPAP Pressure 0 – 15 cm H₂O Continuous Flow 2 – 20 L/min Monitoring Parameters
Parameter Range Inhaled Oxygen Concentration 15% – 100% CPAP Pressure 0 – 60 cm H₂O Continuous Flow 2 – 20 L/min Alarm and Protection Functions
Alarm / Protection Description AC Power Cut-Off Alarm Alerts when AC power is interrupted Low Air / Oxygen Input Pressure Alarm Activates when input pressure is insufficient Air / Oxygen Input Unbalanced Pressure Alarm Warns if input pressures are not balanced High Oxygen Concentration Alarm Triggers if O₂ concentration exceeds safe limits Low Oxygen Concentration Alarm Triggers if O₂ concentration falls below safe limits Pressure Limit Ensures CPAP pressure stays within safe range