top of page
Labmate Single Beam Visible Spectrophotometer LMSV-A100

Labmate Single Beam Visible Spectrophotometer LMSV-A100

Labmate Single Beam Visible Spectrophotometer LMSV-A100 is a high-precision instrument designed for quantitative and photometric analysis across research, clinical, industrial, and educational laboratories. Operating over a wavelength range of 320–1100 nm with a 2 nm spectral bandwidth, it incorporates a high-resolution optical system and imported silicon photodiode detector for rapid, reliable measurements. Equipped with an 11-inch touchscreen and tungsten lamp light source, Labmate LMSV-A100 enables intuitive operation, real-time monitoring, and versatile data output.

  • Product Info

    Key Features

    • Wavelength range: 320–1100 nm with 2 nm spectral bandwidth

    • High-resolution single-beam optical system (Grating 1200 lines/mm)

    • Imported silicon photodiode detector for fast, accurate detection

    • Photometric range: 0–400%T, –4 to +4A with ±0.2%T accuracy

    • 10 mm 4-cell cuvette holder, compatible with multiple cuvette types

    • 11-inch touchscreen display for intuitive operation

    • Output via cloud, email, SMS, Bluetooth, and Wi-Fi

    • Real-time data monitoring with audit tracking and log inquiry

    • Supports thermal and laser printer connections

    Applications

    • Quantitative chemical analysis

    • Nucleic acid and protein assays

    • Pharmaceutical testing

    • Water and environmental pollutant analysis

    • Colorimetric assays and educational laboratory experiments

    Technical Highlights

    Specification Details
    Wavelength Range 320–1100 nm
    Spectral Bandwidth 2 nm
    Optical System Single beam, Grating 1200 lines/mm
    Wavelength Accuracy ±0.1 nm (656.1 nm), ±0.3 nm (other wavelengths)
    Wavelength Repeatability ≤0.1 nm
    Photometric Range 0–400%T, –4 to +4A
    Photometric Accuracy ±0.2%T (0–100%T), ±0.0005A (0–0.5A), ±0.0008A (0.5–1A)
    Photometric Repeatability ≤0.15%T (0–100%T), 0.0004A (0–0.5A), 0.0006A (0.5–1A)
    Resolution 0.00001 Abs
    Stray Light ≤0.03%T at 220 nm, 360 nm
    Stability ±0.0001 A
    Baseline Flatness ±0.0004 A
    Noise 0.0005 A at 500 nm
    Detector Imported Silicon Photodiode
    Light Source Tungsten lamp
    Scanning Speed High/Medium/Low (Max 6000 nm/min)
    Cuvette Holder 10 mm 4-cell holder (compatible with multiple cuvettes)
    Output Cloud, Email, SMS, Bluetooth, Wi-Fi
    Display 11-inch Touchscreen
    Power Supply AC 110–220 V, 50–60 Hz
    Package Dimension 810×660×390 mm
    Gross Weight 27 kg

    Labmate LMSV-A100 combines high accuracy, versatile cuvette compatibility, and advanced data connectivity, making it suitable for rigorous laboratory analysis, industrial testing, and educational applications.