Infitek Fluorescence Spectrophotometer, SP-LF98

Infitek Fluorescence Spectrophotometer, SP-LF98

The Infitek Fluorescence Spectrophotometer SP-LF98 offers precise measurement with an excitation wavelength range of 200nm to 900nm and a wavelength accuracy of ±0.4nm, ensuring reliable results for your applications. Data transfer is streamlined through USB 2.0, enhancing connectivity and ease of use. As an authorized distributor, winselluma provides trusted access to this advanced instrument, backed by dedicated service and support. Explore accurate fluorescence analysis solutions designed to meet professional standards.
  • Product Info

    Fluorescence Spectrophotometer

    SP-LF98 fluorescence spectrophotometer is a new generation of high-performance molecular luminescence analysis equipment. This product is designed for high performance, with high signal to noise ratio, ultra-high scanning speed, ultra-high resolution, ultra-high wavelength accuracy and a variety of accessories. This instrument easily meets the requirements in material research, drug analysis, biochemical and clinical testing, water quality control, food safety testing and other areas of qualitative and quantitative analysis and scientific research. Quantum yield accessory is available for SP-LF98 fluorescence spectrophotometer.

    Main Features:
    1. Horizontal slit optical design, with excellent luminous detection efficiency and high signal to noise ratio. Water Raman peak SNR better than 350: 1 (P-P) or 1000: 1 (RMS). The minimum sample volume is 0.5mL when using a standard 10mm square sample cell.
    2. Wavelength scanning speed up to 60000 nm / min.
    3. Bandwidth 1/2/5/10/20nm adjustable.
    4. Fluorescence Quantum Yield accessory available.
    5. The software offers fluorescence 3D scanning, equal-wavelength difference synchronous scanning, equal wave number difference (constant energy difference) synchronous scanning functions. (synchronous scanning function require pro version software)
    6. Multi accessories available.
    7. Using Hamamatsu’s high quality Xenon light source and photoelectric multiplier tube detectors and instruments to provide sufficient light intensity signal and the detection sensitivity.
    8. Built-in frequency filter. Built-in optical gate, designed for unstable sample.

    SPECIFICATIONS

    Model SP-LF98
    Excitation Source 150W xenon lamp(Hamamatsu)
    Excitation Wavelength 200nm~900nm
    Emission Wavelength 200nm~900nm
    Excitation Slit 1nm/2nm/5nm/10nm/20nm
    Emission Slit 1nm/2nm/5nm/10nm/20nm
    Wavelength Accuracy ±0.4nm
    Wavelength Repeatability ≤0.2nm
    Signal-to-Noise Ratio S/N≥350(P-P)  S/N≥1000(RMS)
    Limit ≤5×10-11 g/ml(Quinine Sulfate Solution)
    Linearity γ≥0.995
    Peak Repeatability ≤1.5%
    Wavelength Scan Speed Multi-speed Level, Maximum at 60000nm/min
    Minimum sample size 0.5mL (Using a standard 10mm square sample cell)
    Octave filter Build in Octave filter
    Photometric Quantity Range -9999 ~ 9999
    Response Time 0.02/0.1/0.5/1/2/4/8 s, Auto adjust
    Data Transportation USB 2.0
    Power 200W
    Power Source 110-240V  AC.50/60Hz
    Instrument Dimension 610×460×365(mm)
    Weight Net Weight:21kg  Gross Weight:26kg